CUE21-A1A, Endress+Hauser, online turbidimeter
CUE21-A1A Endress+Hauser – online turbidimeter. Industrial product for automation, measurement and control.
Cambio 24,43 Kč per 1 EUR. I prezzi sono indicativi, la fatturazione avviene in corone ceche.
Endress+Hauser Turbimax turbidimeters for continuous measurement of water turbidity by ISO 7027 in NTU and FNU. Choose by range and mounting method.
CUE21-A1A Endress+Hauser – online turbidimeter. Industrial product for automation, measurement and control.
I prezzi includono il contributo di riciclaggio.
A turbidimeter monitors turbidity through the optical response of the liquid. It helps capture changes during filtration, water treatment or in a process circuit, but on its own it does not determine the chemical composition of particles. This category offers the Endress Hauser Turbimax CUE21-A1S, online turbidimeter. For selection, the measuring method used and the way a representative sample is supplied are important.
Turbimax CUE21 according to the technical documentation uses an infrared source of 860 nm and measures scattering at an angle of 90° according to the ISO 7027. method. The related CUE22 works with white light. It is therefore not correct to interchange these versions just because of the common name Turbimax.
Particles of different sizes, shapes and optical properties scatter light differently. The same mass of two suspensions may not produce the same turbidity reading. Conversion to mg/l of suspended solids requires a verified correlation for specific samples and conditions. A universal multiplier applicable to all liquids would produce seemingly precise but factually incomparable numbers.
The unit designation must be accompanied by information about the optical method. USGS, in its definition of turbidity, distinguishes infrared and broad-spectrum methods and points out that agreement on a calibration standard does not guarantee agreement on a real sample. Therefore do not automatically convert values from different instruments using the rule 1 NTU = 1 FNU.
When transferring results between operation and the laboratory, record the light source, measuring geometry, unit and sampling method. The documentation of CUE21 also states the NTU/FNU designation; the abbreviation on the display alone must not replace specification of the method. When a specific analytical procedure is required, verify the entire procedure, not just the numerical range of the instrument.
Sampling should preserve the particle distribution in the process. A long tube can retain sediment, a place that is too calm may not represent the flow, and bubbles can create an optical response unrelated to solid particles. The bypass design therefore addresses flow, venting, settling and transport time together. Merely increasing the damping of the record will not remove these problems.
For CUE21, the documentation specifies a sample flow of 0.1–1 l/min. The measuring cuvette is made of borosilicate glass; however, assess chemical compatibility also for the seals, flow head and tubing. The permissible supply pressure and the function of the regulator cannot be confused with an arbitrary direct connection to a process pipe.
When designing the service space, allow access to the cuvette and safe shutdown of the supply. Optical surfaces must remain clean and unscratched; a coating can cause gradual drift that looks like a process change. Plan maintenance according to the medium and check results. Even an automatic cleaning function, if the assembly uses one, does not relieve the operator of the need to verify the condition of the optics.
Store the trend together with service interventions and the state of sampling. A jump after replacing the cuvette or interrupting the flow needs a different explanation than a gradual change during filtration. When comparing with a manual sample, record the exact time and allow for the delay of the bypass branch.
Consider an illustrative check of the same optical method: the reference value is 1.00 FNU and the indication is 1.06 FNU. The difference is +0.06 FNU, relatively 6 %. This result describes only the given check point. It is not the total measurement uncertainty nor a prescribed calibration point or tolerance CUE21. For a decision, the uncertainty of the reference, repeatability, the condition of the cuvette and the rules of the procedure used need to be known.
Dilution can change particle agglomeration and optical conditions. Without a verified procedure, exact proportionality between the result and the dilution ratio cannot be assumed. First assess the appropriate measuring range and representativeness of sampling.
No. Turbidity does not replace microbiological or chemical analysis. Depending on the task, a parallel check in the category may be useful pH measurement, but even a combination of two quantities is not a universal confirmation of quality. For links to other measurements, use the overview Process Sensors.
I cookie tecnici servono per il funzionamento del carrello e dell'ordine — sono sempre utilizzati. I cookie analitici ci mostrano quali pagine le persone cercano e dove si perdono, e ci aiutano a migliorare il negozio. Li utilizziamo solo se ce lo consenti. Maggiori informazioni nella informativa sul trattamento dei dati.
Scegli quali cookie possiamo utilizzare. Puoi modificare la scelta in qualsiasi momento.
Conservano il contenuto del carrello, l'accesso e la lingua scelta e proteggono i moduli dagli abusi. Senza di essi non è possibile completare l'ordine, per questo sono utilizzati anche senza consenso.
Google Analytics ci mostra in forma aggregata il traffico e il comportamento sul sito. Così capiamo quali pagine non funzionano per gli utenti. In base a questi dati non identifichiamo nessuno.